![Chopin Mixolab 2 Maintenance Manual Download Page 62](http://html1.mh-extra.com/html/chopin/mixolab-2/mixolab-2_maintenance-manual_2602981062.webp)
62
Maintenance manual
09/2016
Mixolab 2
Chapter 3 Repairs
1.
Remove the tank
(see op. 1-1)
then empty and dry it.
2.
Remove the 4 retaining screws from the tank base.
3.
Disconnect the connector (
A
) from the detector.
4.
Remove the triangle retaining ring (
B
) and the floater.
5.
Unscrew the detector bracket from the inside using the
MIX-300/OUT tool (
C
).
6.
Take the detector with its coupling washer out of the bowl,
threading the cable through the bottom of the tank.
7.
Take the coupling washer out and discard it.
4: Hydraulic components
Operation 4-1 Removing/replacing the double-level detector with temperature sensor
Page: 1 / 2
Consumables:
Tools:
-
-
A
B
C
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...