52
Maintenance manual
09/2016
Mixolab 2
Chapter 3 Repairs
1.
Remove the Mixolab's left and right-hand covers
(see op. 1-1)
and central cover
(see op. 1-2)
.
Locate each electric wire on the thermocouple's plug.
2.
With a small screwdriver, unscrew the 4 retaining screws on
the thermocouple plug's electric wires (
A
).
3.
Unscrew the thermocouple plug's retaining screw (
B
) then
remove the plug from the frame.
Replacing the thermocouple's miniature plug should not pose any
problems.
2: Bowl mounting plate
Operation 2-4 Removing/replacing the thermocouple's miniature plug
Page: 1 / 1
Consumables:
Tools:
-
Small screwdriver
B
A
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...