![Chopin Mixolab 2 Maintenance Manual Download Page 95](http://html1.mh-extra.com/html/chopin/mixolab-2/mixolab-2_maintenance-manual_2602981095.webp)
95
Maintenance manual
09/2016
Mixolab 2
Chapter 3 Repairs
4.
Calibration window, check that the coefficients CoefEtalTemp1
and 2 = 0 (
C
).
5.
If so, close the window, otherwise set them to 0 as instructed in
the procedure
Modifying the value of the coefficients
.
■
Modifying the value of the coefficients
1.
In the maintenance menu block diagram, untick "Test with
adjusted temperature" then tick it again.
2.
Close the maintenance block diagram. Click the "Calibrate
dough temperature" icon that appears in the main window (
D
).
12: Adjusting and control operations for the Mixolab
Operation 12-4 Calibrating dough temperature
Page: 2/4
Consumables:
Tools:
-
-
C
D
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...