26
Maintenance manual
09/2016
Mixolab 2
Chapter 2 Trouble-shooting
B
Other faults
Symptom/Description
Action
See §
1
Test start-up time very long.
■
Bowl temperature takes too long to fall
(check on the maintenance screen):
- The bowl is cleaned with hot water -> cool
down the bowl before starting a test.
- Problem with cooling circuit (feed water too
hot, insufficient water flow)
-> correct, adjust the flow rate.
- Cooling circuit blocked
-> locate the blockage by dismantling parts
one by one.
■
A sequence of tests with different protocols
in which the tank T° for protocol P1 is different
from that for P2
(Tank T° P1 >> Tank T° P 2).
The tank temperature may be too high (check
on the maintenance screen)
-> empty the tank between two tests
op. 12-7
2
Abnormal change in the unit's
temperature
This phenomenon may be caused by:
■
The cooling circuit's inlet and outlet hoses
being inverted.
■
Cooling water flow incorrectly set -> adjust
■
Check on the magnetic rear plate of the
bowl that the location of the dough
temperature sensor is not blocked up.
op. 12-7
3
Software problem:
Incompatible file formats
■
The format of the test files in the first
software versions is incompatible with that of
the most recent versions
-> to update the Mixolab program, install the
new version in a new directory and keep the
old version.
■
Problem with the pump's solenoid valve.
op. 12-1
4
Abnormal result:
Abnormal start of the torque curve
(very marked peaks during water
injection)
■
Injection nozzle blocked for leaking (on the
sides)
■
Brackets of the mixing blade on the mixing
bowl not fully pushed in, presenting hard
spots.
5
Wear on the bowl's brackets
■
Replace the brackets as soon as traces of
wear appear.
op. 2-7
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...