75
Maintenance manual
09/2016
Mixolab 2
Chapter 3 Repairs
1.
Remove the Mixolab's left and right-hand covers
(see op. 1-1)
and central cover
(see op. 1-2)
.
2.
Remove the wired motor
(see op. 6-1)
.
3.
Loosen the coupling's set screw at the torque meter end (
A
).
4.
Remove the coupling from the torque meter shaft (
B
).
Note: Make sure you retrieve the cotter pin (
C
) on the torque meter
shaft.
To put back the coupling, pay attention to the points below:
- The base of the motor must be parallel to the frame.
- The motor shaft must be aligned with the torque meter and trans-
mission must take place with no hard spots (rotate the thumb wheel
by hand to test this).
- The coupling's screws must be locked with weak Loctite thread-
lock.
- Calibrate the torque meter
(see op. 12-9)
.
6: Motor and transmission
Operation 6-4 Removing/replacing the motor coupling
Page: 1 / 1
Consumables:
Tools:
Plastic clamping ring
Weak threadlock (Loctite)
A
B
C
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...