91
Maintenance manual
09/2016
Mixolab 2
Chapter 3 Repairs
1.
Check on the maintenance screen that the Mixolab's setpoint
temperature for the bowl is preset to 30°C (
A
).
Then let the temperature settle for at least a quarter of an hour.
Note: Take the "full" bowl to measure the temperature of the bowl.
2.
Place the full bowl into its support and lock it in place.
3.
insert the thermometer fully into the opening of the full bowl (
B
).
4.
Wait for 10 minutes then note down the thermometer's temper-
ature reading.
5.
Compare the temperature on the thermometer with the tem-
perature measured by the Mixolab for the bowl (
C
): it should
be the same to within ±0,3°C.
Note: If the temperature is different, go to the maintenance menu
and proceed as follows.
12: Adjusting and control operations for the Mixolab
Operation 12-2 Calibrating the bowl & thermocouple temperature
Page: 1 / 2
Consumables:
Tools:
-
Thermometer with a precision of ±0.1°C
Bowl “full”
A
B
C
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...