![Chopin Mixolab 2 Maintenance Manual Download Page 20](http://html1.mh-extra.com/html/chopin/mixolab-2/mixolab-2_maintenance-manual_2602981020.webp)
20
Maintenance manual
09/2016
Mixolab 2
Chapter 2 Trouble-shooting
ERR 16
"Electronic problem on bowl unit
heater!"
Permanent bowl unit heating
command detected by the heat safety
device
■
Power off then power on the Mixolab.
■
CPU board problem
- Check all electrical connections on the Mix-
olab's CPU board.
- Replace the CPU board if the problem per-
sists.
op. 9-1
ERR 21
"Tank temperature too high! "
Cooling timeout (after 15 minutes' of
no heat, tank T° > 110% of the
setpoint)
■
Confirm on the maintenance screen.
■
The lab temperature is too high (it should
be under the [setpoint T° - 5°C]) -> correct
and wait for the temperature to fall or replace
the water with colder water.
■
Incorrect temperature reading:
- Recalibrate tank temperature
measurement.
- Check and replace the sensor if the
problem persists.
op. 12-3
op. 4-1
ERR 22
"Tank temperature too low!”
Heating timeout (after 15 minutes' of
heat-up, tank T° < 90% of the
setpoint)
■
Confirm on the maintenance screen.
■
Incorrect temperature reading:
- Recalibrate tank temperature
measurement.
- Check and replace the sensor if the
problem persists.
■
Insufficient heating:
- Check the heating circuit
- If necessary, replace the faulty resistors.
op. 12-3
op. 4-1
op. 12-13
op. 2-1
ERR 23
"Tank empty!"
Pictogram at the bottom of the screen
= red
Fill the Mixolab's tank. If there is enough
water:
■
Incorrect reading of water level:
- Sensor floater jammed: dismantle, inspect
and replace if necessary.
- Replace the CPU board if the problem
persists
op. 4-1
op. 9-1
ERR 3
No communication with the Mixolab
when accessing a menu requiring
dialogue
No initialization of the proportioning
pump
■
Check that the Mixolab is connected and
powered up.
■
Check the USB connection between the
PC and the Mixolab (or RS 232 depending
on which one is used), including the internal
wiring.
■
Replace the CPU board if the problem
persists
op. 9-1
ERR no.
Display/Description
Action
See
Chap.3 §
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...