107
Maintenance manual
09/2016
Mixolab 2
Chapter 3 Repairs
Before opening the Mixolab, you must switch it off and dis-
connect its power supply.
1.
Remove the right-hand cover of the Mixolab
(see op. 1-1)
.
2.
Disconnect the connector of the temperature sensor being
tested (see
(A)
)
3.
Measure the sensor's ohmic resistance on the connector plug:
- Bowl unit temperature = between pins [1 or 3 ] and [2 ot 4].
- Tank temperature = between pins [1 or 4 ] and [2 or 5].
4.
The resistance should be 100
:
at 0°C, with a slope of 0.384
:
/°C; i.e. for instance for 25°C: 100 + (0.384x25), namely
109.6
:
.
5.
When you have taken the measurements, put the right-hand
cover of the Mixolab back on
(see op. 1-1)
.
12: Adjusting and control operations for the Mixolab
Operation 12-12 Checking the characteristics of a PT100 temperature sensor
Page: 1 / 1
Consumables:
Tools:
-
Multimeter
A
Bowl temperature
Tank temperature
CPU
board
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...