Chapter 2 Trouble-shooting
ERR 9
"Water dosing problem!"
Pump stroke end timeout when filling;
stroke end contact not reached after
15 seconds
■
Check operation of the pump on the
maintenance screen:
- Place the nozzle on the tank,
- Enter a sufficient water quantity value then
press ON
- Check whether the contact is actuated
("ON") at each rear position of the piston.
■
If the piston does not move or stops
prematurely, there is a mechanical or motor
problem: check the motor's wiring and
connections, and replace the pump If the
problem persists.
■
If the piston correctly moves both ways
and the contact is never actuated ("OFF"),
there is a problem with the contact: check
the contact's wiring and connections, and
replace the pump if the problem persists.
■
If the problem persists, replace the CPU
board.
op. 4-5
op. 4-5
op. 9-1
ERR no.
Display/Description
Action
See
Chap.3 §
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...