Chapter IV Spectrum Analyzer Mode
80
4.3.16 System menu
The System Menu lists the system-related settings of the 4041 series spectrum analyzer. In addition to date and
time, date format, system language, network settings, frequency reference, etc., there are special menus such as
zero frequency calibration and GPS function (option).
The 4041 series spectrum analyzer is featured with zero frequency calibration function, allowing users to perform
zero frequency calibration when needed. When the accurate measurement frequency amplitude of the signal is
lower than 5 MHz, pay attention to the zero frequency signal value. When the zero frequency signal is above -20
dBm, the zero frequency calibration is required to avoid gain compression caused by excessive zero frequency
signal.
The GPS function is an option of the 4041 series. It can be realized through an external GPS antenna. Users can
view the current number of satellites, latitude and longitude, and altitude information.
AliLO Nulling
System
Date
Time
Display
GPS
LAN
>
Freq Ref
Int
Ext
More
1 of 2
>
>
>
>
Ref Out
Off
On
·[AliLO Nulling]:
Press
【
System
】
→[AliLO Nulling] to select zero frequency
calibration when needed by the user, and the calibration result will be saved in the
internal memory of the instrument.
·[Date&Time]:
Press
【
System
】
→[Date&Time] to set the date and time.
·[Display]:
Press
【
System
】
→[Display] to set the display modes, including [Default
Mode], [Black&White], [Night Vision] and other related settings.
·[GPS]:
Press
【
System
】
→[GPS] to pop up GPS-related soft menus, including [GPS
Off On], [GPS Info] and [Reset], so as to set GPS on or off, view GPS details and reset
GPS.
·[Special Attention]
:
The GPS function is an option.
·[LAN]:
Press
【
System
】
→[LAN] to select the network settings of the spectrum
analyzer, including [Prev], [Next], [IP], [Mask], [Gateway] and other soft menus.
·[Freq Ref Int Ext]:
Press
【
System
】
→[Freq Ref Int Ext] to select the internal or
external frequency reference as needed by the user.
·[Ref Out Off On]:
Press
【
System
】
→[Ref Out Off On], and when the frequency
reference is in the internal reference status, choose whether to enable the internal
reference function as needed by the user.
·[Special Attention]
:
The external reference frequency must be 10 MHz±100 Hz
within the amplitude of 0 dBm (limit range: -2 dBm~+10 dBm).
The external
reference frequency must be input from "10 MHz reference input/output" port.
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...