Chapter IV Spectrum Analyzer Mode
50
Fig 4-23 Pulse width displayed by marker
3)
Measurement of pulse repetition frequency (PRF)
The pulse repetition frequency (PRI) is the time interval between any two adjacent pulse responses.
a)
Set the signal generator output signal:
Set the signal generator frequency at 1 GHz and signal power at -20 dBm, and connect the signal generator output
to the input port of spectrum analyzer. Set the pulse modulation to the repetition frequency of 1 kHz, pulse width
of 1 us, and activate the pulse modulation and RF output.
b)
Set the spectrum analyzer:
Press
【
Preset
】
key.
Press
【
Freq
】
, 1 [GHz].
Press
【
Span
】
, 10 [MHz],
【
Sweep
】
, [Sweep Time Auto Man], 1.705 [s].
Press
【
BW
】
, [Res BW Auto Man], 1 [kHz].
Press
【
BW
】
, [Video BW Auto Man], 3 [MHz].
Press
【
BW
】
, [Detector], [Peak] to activate the positive peak detection.
Adjust the span to display the main lobe and at least a pair of sidelobes on the screen.
Readjust the output amplitude of the signal generator to display it on the screen, and reduce the sweep time (i.e.,
quicken the scan speed) until the display is similar as that shown in Fig 4-24.
c)
Measure the pulse repetition interval.
Press
【
Sweep
】
, [Sweep Type Cont Single].
Press
【
Peak
】
,
【
Marker
】
, [Detla],
【
Peak
】
, [Next Pk]. The difference of the two displayed markers equals the
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...