Chapter IV Spectrum Analyzer Mode
30
measurement interface.
f)
Set the center frequency of main channel:
Press [Center Freq], and then set the center frequency of the main channel to 1 GHz with the numeric key.
g)
Set the main channel bandwidth:
Press [Main Ch BW], and then set the bandwidth of main channel to 1 MHz with the numeric key.
h)
Set the adjacent channel bandwidth:
Press [Adj Ch BW], and then set the adjacent channel bandwidth to 2 MHz with the numeric key.
i)
Set the channel offset:
Press [Channel Spacing], and then set the channel offset to 1 MHz with the numeric key.
j)
Start the adjacent channel power ratio test:
Press [ACPR Off On], after that the measurements are displayed at the bottom of the screen. Schematic diagram
of adjacent channel power ratio measurement is as shown in Fig 4-4.
Fig 4-4 Adjacent channel power ratio measurement
k)
Limit setting:
The Limit test function is intended to facilitate the user to observe whether the adjacent channel power exceeds
the set range. Press
【
Measure
】
, [ACPR] and [More 1 of 2] successively to enter the setup menu of adjacent
channel power ratio Limit test;
Press [Upper Limit], and then enter the upper adjacent channel limit with the numeric key;
Press [Lower Limit], and then enter the lower adjacent channel limit with the numeric key.
l)
Enable the limit test function:
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...