Chapter IV Spectrum Analyzer Mode
67
4.3.9
Mode menu
The default function mode in the mode menu is the Spectrum Analyzer mode, which, according to user
requirements, allows the addition of following function options: Interference Analyzer, AM/FM/PM analyzer,
Power meter, Channel scanner and field strength measurement.
中心频率
频率
扫宽
起始频率
终止频率
步进频率
自动
手动
>
Spectrum
Analyzer
Mode
Interference
Analyzer
AM/FM/PM
Analyzer
Power
Meter
Channel
Scanner
Field
Strength
·[Spectrum Analyzer]
: Press [Mode]→[Spectrum Analyzer] to select the Spectrum
Analyzer mode.
·[Interference Analyzer]
: Press [Mode]→[Interference Analyzer] to select the
Interference Analyzer mode. For detailed introduction, see Chapter 5 Interference
Analyzer Measurement Mode.
·[AM/FM/PM analyzer]
: Press [Mode]→[AM/FM/PM analyzer] to select the
AM/FM/PM analyzer mode. For detailed introduction, see Chapter 7 AM/FM/PM
analyzer Measurement Mode.
·[Power meter]
: Press [Mode]→[Power meter] to select the Power meter mode. For
detailed introduction, see Chapter 6 Power meter Mode.
·[Channel Scanner]
: Press [Mode]→[Channel Scanner] to select the Channel scanner
mode. For detailed introduction, see Chapter 8 Channel scanner Measurement Mode.
·[Field Strength Measurement]
: Press [Mode]→[Field Strength] to select the field
strength measurement mode. For detailed introduction, see Chapter 9 Field Strength
Measurement Mode.
4.3.10 Sweep menu
The sweep time specifies the time the spectrum analyzer takes with the local oscillator turned to sweep across a
selected frequency band. The sweep time directly affects the time cost to complete a test. It varies with the span,
RBW and VBW, In
‘
Auto
”
mode, the analyzer maintains the fastest sweep time possible with the selected
measurement settings. The
“
Manual
”
mode can be used to increase the sweep time to satisfy some specific
measurements.
The default sweep mode is linear sweep. The list sweep function is optional, which activates a continuous sweep
that consists of multiple frequency segments. It supports to edit, save and recall the list.
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...