Chapter VIII Channel Scanner Mode (Option)
112
Fig 8-2 Schematic Diagram of the Frequency Scanning
8.1.3
List sweep
Below is an example of the Frequency Scanner mode, mainly involving the following procedures.
1)
Press
【
Sweep
】
→[MScan] → [Edit List] and edit the channel list to be swept. The information of each
channel can be set by setting the signal standard and channel number or setting the frequency and bandwidth
in the list.
2)
Press
【
Sweep
】
→[MScan] →[Number of Channels], and set the number of measured channels. At most 20
channels can be measured at the same time.
3)
Press
【
Sweep
】
→[Display Graph Table] and enable the graph display mode.
4)
Press
【
Sweep
】
→[Power Display Curr Max], and enable the “Current” option. Set the current power of each
channel.