Chapter I Overview
6
Video Bandwidth
1 Hz~10 MHz (1-3 times step)
SSB phase noise
(Carrier 1 GHz, 20
℃
~30
℃
)
≤-102 dBc/Hz@10 kHz frequency offset
≤-106 dBc/Hz@100 kHz frequency offset
≤-111 dBc/Hz@1MHz frequency offset
≤-123 dBc/Hz@10MHz frequency offset
Displayed average noise level
(Tracking Generator off, 50 Ω
input end load, 0 dB input
attenuation, average detection,
video type logarithm, RBW
normalized to 1 Hz, 20
℃
30
℃
)
Preamplifier off
≤
-135 dBm (2 MHz~10 MHz)
≤
-138 dBm (10 MHz~20 GHz)
≤
-135 dBm (20 GHz~32 GHz)
≤
-127 dBm (32 GHz~40 GHz)
≤
-120 dBm (40 GHz~44 GHz)
Preamplifier on
≤
-150 dBm (2 MHz~10 MHz)
≤
-157 dBm (10 MHz~20 GHz)
≤
-154 dBm (20 GHz~32 GHz)
≤
-148 dBm (32 GHz~40 GHz)
≤
-140 dBm (40 GHz~44 GHz)
Second harmonic distortion
(0 dB attenuation, -30 dBm input
signal)
<-60 dBc
Third-order intermodulation
distortion
(-15 dBm double tone signal, 100
kHz interval, 0 dB attenuation,
preamplifier off)
≥+7 dBm 50 MHz~4 GHz
≥+6 dBm 4 GHz~13 GHz
≥+6 dBm 13 GHz~44 GHz
1 dB gain compression
(Dual-tone method test, signal
interval 10 MHz)
≥-2 dBm 50 MHz~4 GHz
≥-3 dBm 4 GHz~13 GHz
≥-3 dBm 13 GHz~44 GHz
Image, multiple and out-of-band
response
(-10 dBm mixer level)
≤-65 dBc 10 MHz~20 GHz
≤-60 dBc 20 GHz~44 GHz
Residual response
(RF input matching, 0 dB
attenuation, Tracking Generator
off)
Exceptional frequency point 3,200 MHz
Preamplifier on:
≤-100 dBm 10 MHz~20 GHz
≤-95 dBm 20 GHz~44 GHz
Preamplifier off:
≤-90 dBm 10 MHz~13 GHz
≤-85 dBm 13 GHz~20 GHz
≤-80 dBm 20 GHz~44 GHz
Scale fidelity
±1.00 dB
Total level uncertainty
±1.8 dB (10 MHz~13 GHz)
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...