Chapter IV Spectrum Analyzer Mode
53
a)
Press
【
Freq
】→
[Start Freq], and set the start frequency to 100 MHz;
b)
Press
【
Freq
】→
[Stop Freq], and set the start frequency to 4 GHz;
c)
Press
【
Measure
】→
[Generator]
→
[Generator Off On] to turn on the generator switch;
d)
Press
【
Measure
】→
[Generator]
→
[Power -20 dBm] to set the power of output signal.
e)
Press
【
Measure
】→
[Generator]
→
[Mode CW Track] to switch the signal generator mode to the tracking mode.
The test result is shown in Fig 4-26 below:
Fig 4-26 Test screen of tracking generator
3
)
Normalization measurement of tracking generator
Normalization measurement reduces the cable loss during the measurement process. It can reflect the
amplitude-frequency characteristic of tested object in a more accurate way. For example of testing the
amplitude-frequency characteristic of 2.3 GHz~2.4 GHz band-pass filter, refer to the following steps:
a)
Press
【
Freq
】→
[Start Freq], and set the start frequency to 2.1 GHz;
b)
Press
【
Freq
】→
[Stop Freq], and set the start frequency to 2.5 GHz;
c)
Press
【
Measure
】→
[Generator]
→
[Generator Off On] to turn on the generator switch;
d)
Press
【
Measure
】
→
[Generator]
→
[Mode CW Track] to switch the signal generator mode to the tracking mode;
e)
Press
【
Measure
】→
[Generator]
→
[Trans Meas]
→
[Normalize Off On] at the cable connection signal generator
RF output end and spectrum RF input end to turn on the normalization switch;
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...