Appendix A Test Results of Performance Characteristics
190
Table A.2
4041 Series Spectrum Analyzer Record Sheet
(Tracking Generator Option)
S/N
Test Item
Unit
Standard Requirements
Test Result
1
Frequency range
/
Lower limit of the frequency range:
100 kHz
/
Upper limit of the frequency range:
20 GHz
2
Frequency
accuracy
kHz
1 GHz
±0.80
kHz
2 GHz
±1.60
kHz
6 GHz
±4.80
kHz
18 GHz
±14.40
3
Phase noise
(6GHz carrier)
dBc/Hz
+10 kHz
-85
dBc/Hz
-10 kHz
-85
dBc/Hz
+100 kHz
-90
dBc/Hz
-100 kHz
-90
4
Amplitude range
/
Range: -5 dBm~-35 dBm
5
Amplitude
accuracy
dB
100 MHz, -5 dBm
output
±3.5
dB
100 MHz, -10 dBm
output
±3.5
dB
100 MHz, -20 dBm
output
±3.5
dB
100 MHz, -30 dBm
output
±3.5
dB
100 MHz, -35 dBm
output
±3.5
dB
4 GHz, -5 dBm output
±3.5
dB
4 GHz, -10 dBm output
±3.5
dB
4 GHz, -20 dBm output
±3.5
dB
4 GHz, -30 dBm output
±3.5
dB
4 GHz, -35 dBm output
±3.5
dB
8 GHz, -5 dBm output
±3.5
dB
8 GHz, -10 dBm output
±3.5
dB
8 GHz, -20 dBm output
±3.5
dB
8 GHz, -30 dBm output
±3.5
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...