Chapter XI Performance Characteristics Test
157
26) Set the frequency of the calibration factor of the power meter to 15 GHz.
27) Disable the difference marker function of the spectrum analyzer and set the frequency to 15 GHz, with other
settings left unchanged.
28) Repeat steps 6~16.
11.11 1dB gain compression
Description:
When the input signal level increases, the unit circuits of such devices as spectrum analyzer mixer,
amplifier, etc., can work near the saturation point. At this time, the output signal component no longer changes
linearly with the input signal. Usually, the input level corresponding to output 1dB lower than linear value is used
to represent 1 dB compression point.
This test measures the gain compression of the spectrum analyzer with two signals spaced by 10 MHz. During the
test, input a small signal to the spectrum analyzer (below -10 dBm) at first. Then input the specified large
amplitude signal into the spectrum analyzer. The reduction in the amplitude of the first signal caused by the
second signal is the measured gain compression.
A) Testing equipment
Power meter ............................................................................................................................................... ML2437A
Power sensor .............................................................................................................................................. MA2445D
Synthesized signal generator#1 ....................................................................................................................... 1464C
Synthetic signal generator #2 .............................................................................................................................. 1461
RF directional coupler ...................................................................................................................................... 70607
Millimeter wave directional coupler ................................................................................................................. 70603
b) Adapter
2.4 mm (f)- 2.4 mm (f) adapter * 2
BNC T type (m) (f) (f) adapter * 1
2.4 mm (f)-3.5 mm (m) adapter * 2
c) Cable
BNC (m-m) cable * 2
2.4 mm (m-m) cable * 2
d) Test steps
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...