Chapter IV Spectrum Analyzer Mode
49
Press
【
Peak
】
,
【
Marker
】
, [Detla],
【
Peak
】
, [Next Pk].
The amplitude difference between the main lobe and sidelobe is the sidelobe ratio, as shown in Fig 4-22.
Fig 4-22 Sidelobe ratio displayed beside the marker
e)
Measure the pulse width, which equals the reciprocal of frequency difference between the two sidelobe
envelope peaks:
Press
【
Marker
】
, [Detla],
【
Peak
】
, [Next Pk Right], [Next Pk Right].
At this time, the reciprocal of frequency difference of the read differential marker is the pulse width, as shown in
Fig 4-23. In order to obtain the most accurate pulse width, adjust the marker position manually to measure the
distance between zero crossing points of two adjacent sidelobes. If the resolution bandwidth is reduced to make
the zero crossing point more sharp, the measurement precision will be higher.