![Ceyear 4041 Series User Manual Download Page 145](http://html1.mh-extra.com/html/ceyear/4041-series/4041-series_user-manual_2584923145.webp)
Chapter X Working Principle
133
FPGA after the AD conversion to complete the digital down-conversion, IF filtering. video filtering, etc.
CPU board consists of the power conversion and management circuit, interface and display circuit, FPGA/CPLD
control and digital signal processing circuit, and ARM core module to realize the AC/DC conversion of power
supply, power supply management, digital signal processing, logic circuit control, etc., and outputs the test results
to the liquid crystal display in the form of sweep curve, waterfall map, etc.
Frequency synthesizer board can provide 10MHz reference input/output function externally, and the internal
reference loop output signal serves as the frequency reference of the local oscillation and provides the
high-stability clock signals for AD sampling and FPGA. Local oscillation generation circuit provides the local
oscillation signal for the mixer in the microwave and millimeter wave frequency conversion module, which is
related to the hardware sweep speed, phase noise, sideband stray and other specifications of the whole machine.
Tracking generator board (option) can output the dot frequency signal with the fixed frequency and power and
also the fixed power, and set the frequency sweep signal in the frequency band.
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...