![Ceyear 4041 Series User Manual Download Page 167](http://html1.mh-extra.com/html/ceyear/4041-series/4041-series_user-manual_2584923167.webp)
Chapter XI Performance Characteristics Test
155
coupler directly to the RF input port of the spectrum analyzer.
9)
Press
【
Peak
】
on the spectrum analyzer and wait for completion of a new sweep. Press
【
Marker
】
[Detla],
【
Freq
】
【
↑
】
.
10) Turn on the RF output of the synthesized signal generator #2, and set the power to –15 dBm.
11) Press
【
Peak
】
on the spectrum analyzer.
12) Adjust the power level of synthesized signal generator #2 so that the differential marker reading of the
spectrum analyzer approaches 0 dB.
13) Press
【
Marker
】
[Detla],
【
Freq
】
【
↓
】
【
↓
】
on spectrum analyzer. Wait for completion of new sweep.
Press
【
BW
】
[Average On] 5 [Enter]. Wait for the video is averaged 5 times, and press
【
Peak
】
.
14) The peak-marker reading of spectrum analyzer is the low-end third-order intermodulation distortion product
△
A. According to the third-order intermodulation distortion product, the third-order intercept point is
calculated as follows (where L
0
is the level of mixer):
0
2
A
TOI
L
(6)
15) On the spectrum analyzer, press
【
Freq
】
【
↑
】
【
↑
】
【
↑
】
, wait for completion of new sweep, and then
press
【
Peak
】
.
16) The marker difference at this time is the high-end intermodulation distortion product. Calculate the
third-order intercept point according to Equation 6 and compare the high-end intermodulation distortion
product with the low-end intermodulation distortion product. Record the smaller value as the final test result
in the corresponding test item in the “4041 Series Spectrum Analyzer Record Sheet”.
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...