Chapter I Overview
7
(after 30 minute warm-up,input
signal 0 dBm ~-50 dBm, all set as
auto coupling,20
℃
~30
℃
)
±2.3 dB (13 GHz ~40 GHz)
Input attenuator
Attenuation range: 0 dB~50 dB, 10 dB step
Switching uncertainty: ±1.20 dB
Maximum safe input level
(
CW input
)
+30dBm, typical value (
≥
10dB attenuation, preamplifier off )
+23dBm, typical value (<10dB attenuation, preamplifier off )
+13dBm, typical value (preamplifier on)
Reference Level
Logarithmic scale: -120 dBm~+30 dBm, 1 dB step
Linear scale: 22.36 uV~7.07 V, 0.1% step
Switching uncertainty: ±1.20 dB (reference level: 0 dBm~-60 dBm)
Video Bandwidth
1 Hz~10 MHz (1-3 times step)
Input voltage standing-wave ratio
(>10 dB input attenuation)
≤ 1.80:1 50 MHz~20 GHz
≤ 2.20:1 20 GHz~44 GHz
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...