Chapter VIII Channel Scanner Mode (Option)
111
Fig. 8-1 Schematic Diagram of Channel Scanner
8.1.2
Frequency Scanner
Below is an example of the Frequency Scanner mode, mainly involving the following procedures.
1)
Press
【
Sweep
】
→ [FScan] → [Start Freq], and set the center frequency of the starting channel.
2)
Press
【
Sweep
】
→ [FScan]→ [Step Freq], and set the frequency step of the measured channel.
3)
Press
【
Sweep
】
→ [FScan] → [Bandwidth], and set the bandwidth of the measured channel.
4)
Press
【
Sweep
】
→ [FScan] → [Number of Channels], and set the number of measured channels. At most 20
channels can be measured at the same time.
5)
Press
【
Measure
】
→ [Display Graph Table] and enable the graph display mode.
6)
Press
【
Measure
】
→ [Power Display Curr Max], and enable the “Current” option. Set the current power of
each channel.
7)
Press
【
Measure
】
→ [Color Code Single Dual] and enable the dual color mode.
8)
Press
【
Measure
】
→ [Orientation Vert Hori] and enable the horizontal mode.
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...