Chapter IV Spectrum Analyzer Mode
28
g)
Select the measurement method:
Press
【
Measure
】
, [OBW] and [Method % dBc] successively, and then select the measurement method of
occupied bandwidth. The measurement method can be set to the power percentage method or the dBc down
method. The underlined mode is the mode selected currently, and the percentage method is set as default.
Fig 4-3 Occupied bandwidth measurement
h)
Change the percentage:
If you decide to use the power percentage method, press
【
Measure
】
, [OBW] and [% 99%] successively and
then change the percentage with the numeric key, the Up or Down key or the knob. The set value of percentage
ranges from 10% to 99.99% with a minimum step of 0.01%, and 99% is set as default.
i)
Change the value of dBc:
If you decide to use the dBc down method, press
【
Measure
】
, [OBW] and [dBc -3.00 dB] successively and then
change the value of X dB with the numeric key, the UP or DOWN key or the knob. The set value of X dB ranges
from -0.1 dB to -100 dB with a minimum step of 0.01 dB, and -3 dB is set as default.
j)
Change the occupied bandwidth span:
Press
【
Measure
】
, [OBW] and [Span] successively, enter the occupied bandwidth span with the numeric key, and
then enter the unit with the soft key. The default is 3 MHz.
k)
Disable the occupied bandwidth measurement:
Press
【
Measure
】
, [OBW] and [OBW Off On] successively, and then select Off to exit occupied bandwidth
measurement. After that, the analyzer will switch back to the spectrum measurement interface.
4.1.3
Adjacent channel Power meter
In this section, the adjacent channel power ratio measurement of FM signal is taken as an example to demonstrate
the method to measure the adjacent channel power ratio with the 4041 series spectrum analyzer.
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...