Chapter IV Spectrum Analyzer Mode
42
Fig 4-14 Input signal and harmonic wave
c)
Set the video bandwidth to smooth the noise in order to increase the resolution.
Press
【
BW
】
, [Video BW Auto Man] to activate the manual function.
Reduce the video bandwidth by the key
【
↓
】
to decrease the video bandwidth.
d)
In order to improve the measurement precision, set the fundamental wave peak level to the reference level.
Press
【
Peak
】
, [Peak Search], and read the power of peak.
Press
【
Ampt
】
, [Ref Level], and set it to the power of peak, as shown in Fig 4-15.
Fig 4-15 Setting the signal peak to the reference level to obtain maximum precision