Chapter IV Spectrum Analyzer Mode
40
e)
Set the attenuator to 0 dB.
Press
【
Ampt
】
, [Atten Auto Man], 0 [dB].
Press
【
Peak
】
, [Peak Track Off On].
Press
【
Marker
】
, [Detla].
Fig 4-12 Observation of harmonic distortion
f)
Increase the attenuator setting to 10 dB;
Press
【
Ampt
】
, [Atten Auto Man], 10 [dB].
Note to observe the reading of differential marker, as shown in Fig 4-13. This reading is the distortion difference
generated by the attenuator set at 0 dB and 10 dB. When the attenuator setting is changed, if the difference
frequency marker reading is 1 dB or greater, it means the spectrum analyzer has a certain degree of distortion.
When the differential marker reading is not obvious, the attenuation can be increased.