Chapter IV Spectrum Analyzer Mode
33
Fig 4-7 Place the signal peak at the reference level
f)
Set the second marker and measure the distortion product:
Once the marker is activated, the differential marker function can generate the second marker and display the
difference of the two markers. It is convenient to perform relative measurement at this time.
Press
【
Peak
】
to activate a marker.
Press
【
Marker
】
and [Detla] to activate the second marker.
Press
【
Peak
】
and [Next Pk Left] or [Next Pk Right] to set the second marker at the peak point of distortion
product near the signal of signal generator. As shown in Fig 4-8, the frequency and amplitude differences of the
two markers are displayed at the marker display area, and the amplitude difference of the markers is the
measurement of third-order intermodulation distortion.
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...