Chapter XI Performance Characteristics Test
161
11.13 Residual response
Description:
The residual response is the discrete response observed on the display when the spectrum analyzer
doesn’t receive input signals.
a) Test device
50
Matched load
b) Test steps
Fig 11-14 Residual Response Testing Setup
Residual response, 10 MHz~20 GHz (preamplifier on)
1)
Connect the test device according to Fig 11-14. Connect the 50
matcher with the signal input port of the
spectrum analyzer, and set the spectrum analyzer as follows:
Center frequency .................................................................................................................................. 65 MHz
Frequency step ................................................................................................................................... 100 MHz
Span ................................................................................................................................................... 110 MHz
Reference level ................................................................................................................................... -50 dBm
Preamplifier ................................................................................................................................................. On
Resolution bandwidth ........................................................................................................................... 10 kHz
Video bandwidth ..................................................................................................................................... 3 kHz
2)
Activate the limit line function and set the upper limit line at -100 dbm. The noise level shall be at least 5 dB
below the limit line. If not, it is necessary to reduce the noise level by reducing the resolution BW and video
BW.
3)
Observe whether there is residual response signal on the noise baseline of the spectrum analyzer. If there is
residual response signal, read the residual response amplitude with the marker and record the measurement
results. During the test, make the residual response signal amplitude 10 dB higher than the displayed average
noise level at least. If the residual response signal amplitude is smaller, further reduce the resolution BW to
reduce the displayed average noise level.
4)
Change the center frequency according to [Frequency [Center Freq]
【
↑
】
. Repeat step 3 Check the residual
response of the frequency to 20 GHz, and record the tested maximum response value in the measurement
result of the corresponding test item in the
4041 Series Spectrum Analyzer Record Sheet
.
Residual response, 20 GHz~44 GHz (preamplifier on)
50
Matcher
Spectrum
Analyzer
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...