Chapter V Interference Analyzer Mode (Option)
88
5.3.2
Span menu
Span
Full Span
Zero Span
Last Span
< Back
Span
·[Span]:
Press
【
Span
】
→[Span] to activate the span submenu. You can change it by
using the numeric keys and selecting a frequency unit, or by using
【
↑
】
【
↓
】
keys and
the knob. Refer to the description of [Span] menu for specific introduction.
·[Special Attention]: Change the span by using
【
↑
】
【
↓
】
keys and the knob based
on the steps of 1-2-5.
·[Full Span]:
Press
【
Span
】
→[Span]→[Full Span] to set the current span to 44.1 GHz.
·[Span]:
Press
【
Span
】
→[Span]→[Zero Span] to set the current span to 0 Hz.
·[Last Span]:
Press
【
Span
】
→[Span]→[Zero Span] to restore the span to that last set.
·[Back]:
Press
【
Span
】
→[Span]→[Back] to return to the
【
Freq
】
menu.
·[Special Attention]: It can only be zero span under the RSSI mode.
5.3.3
Amplitude menu
Ref Level
0.0dBm
Amplitude
Ref Position
Atten
Auto Man
Scale/Div
20.0dB
Units
dBm
Pre Amp
Off
On
·[Ref Level]:
Press
【
Ampt
】
→[Ref Level] and select [dBm], [-dBm], [mV] or [uV] in
the frequency unit menu by using the numeric keys on the front panel, or
【
↑
】
【
↓
】
keys and the knob.
·[Special Attention]: The step is 10 dB when the
【
↑
】
【
↓
】
keys and the knob are
used.
·[Ref Position]:
Press
【
Ampt
】
→[Ref Position] and change it by using the numeric
keys or by using
【
↑
】
【
↓
】
keys and the knob.
·[Atten Auto Man]:
Press
【
Ampt
】
→[Atten Auto Man] and switch the attenuator auto
or manual mode in the menu by using the numeric keys or by using
【
↑
】
【
↓
】
keys
and the knob.
·[Special Attention]: The setting range of the attenuation value is between 0 dB
and 60 dB, and the step is 10 dB.
·[Scale/Div]:
Press
【
Ampt
】
→[Scale/Div] and set it between 0.1 dB and 20 dB by
using the numeric keys or by using
【
↑
】
【
↓
】
keys and the knob.
·[Units]:
The amplitude in the Interference Analyzer mode is in dBm.
·[Pre Amp Off On]:
Press
【
Ampt
】
→[Preamplifier Off On] to turn the preamplifier on
or off.
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...