![Ceyear 4041 Series User Manual Download Page 192](http://html1.mh-extra.com/html/ceyear/4041-series/4041-series_user-manual_2584923192.webp)
Appendix A Test Results of Performance Characteristics
180
Table A.1 (Continued 2) 4041 Series Spectrum Analyzer Record Sheet
S/N
Test Item
Unit
Standard Requirements
Test Result
8
Resolution BW
accuracy
/
10 kHz
±10.0%
/
3 kHz
±10.0%
/
1 kHz
±10.0%
9
Resolution
Bandwidth
Switching
uncertainty
dB
10 MHz
±1.20
dB
3 MHz
±1.20
dB
1 MHz
±1.20
dB
300 kHz
±1.20
/
100 kHz
Reference
dB
30 kHz
±1.20
dB
10 kHz
±1.20
dB
3 kHz
±1.20
dB
1 kHz
±1.20
dB
300 Hz
±1.20
dB
100 Hz
±1.20
dB
30 Hz
±1.20
dB
10 Hz
±1.20
dB
3 Hz
±1.20
dB
1 Hz
±1.20
10
Sideband noise
(carrier frequency
1GHz)
dBc/Hz +10 kHz
≤
-102
dBc/Hz -10 kHz
≤
-102
dBc/Hz +100 kHz
≤
-106
dBc/Hz -100 kHz
≤
-106
dBc/Hz +1 MHz
≤
-111
dBc/Hz -1 MHz
≤
-111
dBc/Hz +10 MHz
≤
-123
dBc/Hz -10 MHz
≤
-123
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...