Chapter XI Performance Characteristics Test
160
2.4 mm (f)-2.4 mm (f) adapters * 2
c) Cable
BNC (m-m) cable * 1
2.4 mm (m-m) cable * 1
d) Test steps
Fig 11-13 Image, Multiple, and Out-of-band Response Testing setup
1)
Connect the test instrument as shown in Fig 11-13. The spectrum analyzer provides a reference frequency
for the synthesized signal source.
2)
Sets the frequency of 1464C to 2 GHz, and power to -10 dBm.
3)
Set up the spectrum analyzer as follows:
Center frequency .......................................................................................................................................... 2 GHz
Span ............................................................................................................................................................ 10 kHz
Reference level ......................................................................................................................................... -10 dBm
Resolution bandwidth ................................................................................................................................... 10 Hz
Video bandwidth ........................................................................................................................................... 10 Hz
4)
Adjust the output power level of synthesized signal generator so that the peak value of the signal close to the
reference level of the spectrum analyzer.
5)
Press
【
Peak
】
【
Marker
】
[Detla] on the spectrum analyzer.
6)
Set the 1464C to each frequency point where image, multi and out-of-band responses can occur at 2GHz in
the corresponding test item listed in the
“
4041 Series Spectrum Analyzer Record Sheet
”
. Set the reference
level of spectrum analyzer to -40 dBm and the resolution bandwidth to10Hz. Press
【
Peak
】
, and record the
differential marker amplitude as the response amplitude in the corresponding test item of the
“
4041 Series
Spectrum Analyzer Record Sheet.
7)
For the image, multiple, and out-of-band responses of the remaining frequency points listed in the "4041
Series Spectrum Analyzer Record Sheet", refer to steps 2 to 6 for testing until all frequency points have been
tested.
BNC cable
Synthesized
signal generator
Spectrum Analyzer
Adapter
10 MHz IN
2.4 mm cable
Adapter
10 MHz output
Summary of Contents for 4041 Series
Page 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ...8 Article I Instructions...
Page 21: ...9...
Page 33: ...Chapter III Basic Operations 21...
Page 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Page 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Page 142: ...130 Article II Technical Instructions...
Page 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Page 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Page 185: ...173 Article III Maintenance Instructions...
Page 186: ...174...