System Error Codes | 285
Code
Message
What to do
217
Strip washer syringe motor
didn’t find home opto sensor
transition
Run self test. If error reoccurs, contact
BioTek TAC.
218, 228 Aspiration head motor didn’t
find home opto sensor
transitionAspiration head motor
didn’t find autocal jig opto
sensor transition
Run self test. If error reoccurs, contact
BioTek TAC.
219, 229 RAD Y-axis motor didn’t find
home opto sensor transitionRAD
Y-axis motor didn’t find autocal
jig opto sensor transition
Run self test. If error reoccurs, contact
BioTek TAC.
220
Carrier X motor didn’t find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.
221
Carrier Y motor didn’t find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.
300
Carrier X motor interlock safety
switch open
Service Only.
Contact BioTek TAC.
301
Carrier Y motor interlock safety
switch open
Service Only.
Contact BioTek TAC.
302
Dispense head motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
303
Secondary Peri-pump motor
interlock safety switch open
Service Only.
Contact BioTek TAC.
304
Syringe A motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
305
Syringe B motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
306
Peri-pump Pump Cover is open
Close the pump cover door and rerun the
protocol.
307
Strip washer motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
308
Aspirate head motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
BioTek Instruments, Inc.