Perform the Syringe Dispense Precision & Accuracy Test | 240
12. Analyze your test results. The following is the Pass criteria for each test:
Test
%CV
% Accuracy Error
Test 1
< 6.0%
± 5.0%
Test 2
< 3.0%
± 1.25%
Test 3
< 12.0% ± 5.0%
6 Well
< 6.0%
± 1.0%
All other plate-type tests
< 4.0%
± 1.0%
If one or more of your tests are failing, clean the dispense tubes with the stylus, re-
prime the manifold, and repeat the test(s). If your tests continue to fail, contact
BioTek’s Technical Assistance Center.
Documenting Test Results
Dispense Precision & Accuracy Test Worksheets are provided on the MultiFlo FX
Operator's Manual USB flash drive in the Qualification chapter PDF. We
recommend you make copies of the appropriate pages and use them to record your
calculations and test results.
Each worksheet records calculations and pass/fail test results for an individual test.