158 | Chapter 3: Operation
Require a Specific Peri-pump Cassette
When defining a Peri-pump step you can require a specific cassette type. Choose
the desired behavior when the required cassette is
not
installed at runtime:
l
Prompt
the user to confirm
: provides the best protection against an unintentional
mismatch. If the cassette type itself has been physically changed to match the protocol,
but the instrument's setting has not been updated, this option changes the MultiFlo FX's
setting upon confirmation. However, if the cassette has not been changed to match the
protocol, users are given a chance to cancel the run, fix the error, and rerun the
protocol.
l
Return an
Error
code
: gives robotics programmers the ability to design and run
unattended processing routines without fear of a message screen interrupting the
operation.
l
Automatically
Set
the cassette type
: this option changes the MultiFlo FX's setting
without a confirmation. This option is for advanced users only.
n
No action is taken when the cassette type setting matches the protocol's
required cassette type.
n
RAD technology cassettes add a layer of complexity to this feature. Error code
6029 is displayed when a mismatch of required or installed cassette types
occurs. Be as precise as possible when using this feature for best results.
Note:
The
Cassette Requirement Mode
is automatically enforced when a 0.5 µL
dispense is requested. In this case, make sure a 1 µL cassette is installed.
LHC
Touch screen
1. Select
Tools>Instrument Utilities>Peri-
pump Dispenser
.
2. Select the desired
Cassette Requirement
Mode
behavior:
o
PROMPT
o
ERROR
o
SET
1. From the home screen,
select
Instrument>Peri-
pump
.
2. Select the
Cassette
Requirement Mode
option:
o
PROMPT
o
ERROR
o
SET
MultiFlo™ FX Multi-Mode Dispenser