Peri-pump Dispenser Maintenance | 195
Procedure
1. Fill the 10 cc syringe with water and set aside.
2. Remove the cassette from the dispenser.
3. Use the screwdriver to open the
Tip Holder
. Put the top of the holder aside.
4. Lift the affected dispense tube from the holder and pull its tip off the tube.
5. Slide the tip, tapered end first, into the tubing on the end of the syringe.
6. With the tip poised to expel the clog and the water into the beaker or a sink,
discharge the syringe.
7. Fill and discharge the syringe as many times as needed to flush the tip.
8. Reassemble the cassette:
l
Push the straight end of the tip into the bottom of the tube (the tapered end of the
dispense tip is exposed).
l
Reinsert the tube into the Tip Holder. Seat the flared edges of the tip into the molded
slots.
l
Replace the Tip Holder cover with its two screws. The etched BioTek label identifies the
top of the cover (except for 1536 cassettes’ steel cover plate).
RAD™ Technology Cassettes
Gaining access to the tips is the only difference in the procedure above for
unclogging the single-tube and 8-to-1 bulk-dispensing cassettes. Here are some
guidelines for disassembling and reassembling the RAD cassettes:
l
See RAD Single-tube Cassette - Unclog the Tip on the next page
l
See RAD 8-to-1 Cassette - Unclog the Tips on page 197
BioTek Instruments, Inc.