84 | Chapter 2: Installation
For these
MultiFlo FX
models:
l
MFX
l
MFXP
First remove the
plate carrier
shipping bracket
and then remove
the dispense arm
bracket.
Reverse this
order when
reinstalling the
brackets (before
shipping): put the
dispense arm
bracket on the
post first.
Install shipping brackets:
1. First, remove the dispense cassette, Syringe dispenser manifolds or spacer block
and washer manifolds, as needed.
2. Lower the aspiration arm as far as it will go and raise the dispense arm. (Does
not apply to MFX and MFXP models.)
3. Secure the dispense arm with its shipping bracket, aligning it with its screw
holes, and use the Allen wrench to install the two screws that hold it to the
instrument.
4. Slide the plate carrier to the left, positioning it below the dispense arm.
5. Insert the plate carrier shipping bracket: rest the bracket on its post and align it
with the screw holes; use the Allen wrench to insert the two screws to hold it to
the plate carrier.
With the shipping hardware installed, you are ready to pack the instrument in its
shipping container.
MultiFlo™ FX Multi-Mode Dispenser