168 | Chapter 3: Operation
Autoclavable
Non-autoclavable
Stainless steel and glass pump heads
Black plastic pump heads
Glass supply bottles
Plastic supply bottles
Amber, transparent check valves
White, opaque check valves
Changing the Instrument's Settings
About the Onboard Settings
The instrument's onboard settings dictate its behavior. Certain settings, like Plate
Clearance, can improve performance. Other settings are critical to its performance.
For example, the manifold setting must match the installed manifold type.
LHC users: do not confuse the LHC's
Target Instrument Settings
assigned to each
protocol with the onboard settings! Review the Help topic: LHC Protocols Explained to learn
the distinctions.
To change the onboard settings:
LHC
Touch screen
1. Select
Tools>Instrument
Utilities
2. Choose the applicable tab and
modify settings as desired.
3. Click
Send
after changing a
setting to update the
instrument.
1. Press
Instrument>Options
for washer settings.
2. Explore the other
tabs for more
settings.
n
When using the LHC to change instrument settings, you may need to reboot the
MultiFlo FX
before
controlling it with the touch screen.
Dispense Pattern
Touch screen: Select Instrument>General
Select the fill pattern: by
column
or
row
for processing 384- and 1536-well plates.
For high-density plates, the 8-tip manifold must address the plate multiple times to
fill it. Column-wise dispensing fills each column before moving to the next. Row-
MultiFlo™ FX Multi-Mode Dispenser