Decontamination | 205
3. Use the hex wrench to loosen the setscrew on top of the sleeve that holds the
piston and then remove the piston.
Important!
Autoclave the piston and syringe head separated from one another.
Keep the piston and syringe head unattached to each other when autoclaving.
4. Autoclave at 134°C and 216 kPa for 3 minutes, or 121°C and 115 kPa for 30
minutes. The manifold, tubing, autoclavable check valves, and supply bottles
may also be sterilized in the autoclave.
n
Check valves (PN 68073) recommended for use with organic substances
cannot
be autoclaved.
5. Replace the components by reversing the steps:
6. With the flat side of the shaft facing up, slide the syringe piston shaft into the
piston holder until it stops.
7. Use the 3/32” (2.39 mm) hex wrench to tighten the setscrew.
8. Push the syringe head over the piston until it is flush with the unit and use the
hex wrench to attach the two mounting screws.
Decontamination
Any laboratory instrument that has been used for research or clinical analysis is
considered a biohazard and requires decontamination prior to handling.
Decontamination minimizes the risk to all who come into contact with the
instrument during shipping, handling, and servicing. Decontamination is required
by the U.S. Department of Transportation regulations. Persons performing the
decontamination process must be familiar with the basic setup and operation of the
instrument.
BioTek Instruments, Inc.