
MultiFlo FX Software Error Codes | 297
Error
Code
Description
Help
6044
Serial port error
Fix the COM port setting. Check the
cabling. Click the Port link and use the
drop-down menu to see all active ports.
Customize the Predefined Protocols to
avoid this error in future.
When controlling the BioStack with the
LHC, make sure the instrument's
BioStack setting is
Manual
.
6045
Serial write error
6046
Serial read error
6047
Checksum error
Contact BioTek TAC.
6048
Serial NAK error
Make sure the COM port setting is correct
and the cable is properly connected.
Restart the instrument. If error reoccurs,
contact BioTek TAC.
6049
Excess data, or not enough data,
received.
To correct these errors:
l
Reset the instrument.
l
Check cables, plug in only one
communication cable at a time:
USB or serial.
l
Try running a different protocol.
If error reoccurs, contact BioTek TAC.
6050
Invalid message header
6051
Invalid message object
6052
Invalid message body size
6053
Serial message timeout
6054
Port handle error
6055
Read timeout value is invalid.
6056
Unauthorized to open the COM port.
Make sure the COM port setting is correct
and the cable is properly connected.
Restart the instrument. If error reoccurs,
contact BioTek TAC.
6057
Out-of-range parameter for the open
port function.
6058
Unable to open the COM port.
6059
Unable to clear the transmission
buffer.
6060
Unable to close the port.
6061
Port is no longer available.
6062
Unhandled exception while
transmitting message.
Contact BioTek TAC
6063
The selected plate type is not allowed
with this protocol step.
Modify the protocol to change the plate
type.
BioTek Instruments, Inc.