Overnight/Multi-Day Maintenance | 179
Overnight/Multi-Day Maintenance
Overnight/multi-day maintenance involves flushing all solutions out of the
instrument, and then periodically rinsing and soaking the tubes to keep them
moist. Here are three recommendations for accomplishing the task. Employ the
method that best suits your work flow:
Overnight Loop
To keep the wash manifolds in a wetted condition, you can run these predefined
protocols to soak the tubes for several hours at a time:
l
W-OVERNIGHT_LOOP
: requires the washer to remain turned on.
l
W-RINSE_AND_SOAK
: alternatively, run this protocol and turn off the
instrument after the soak begins. The tubes will soak in the priming trough
until the instrument is turned on again.
AutoPrime
Recommended for optimum performance, AutoPrime keeps the tubing wet in
between runs and can be an essential part of your daily maintenance routine.
About AutoPrime
AutoPrime
automatically primes the tubing whenever the instrument is idle for a
specified time. Keeping the tubes wet prevents clogging and mitigates fluid
evaporation at the tips. AutoPrime's submerge feature lets you soak the tubes for
extended periods, which is an effective maintenance option.
The downtime interval is defined for all components, but each device supports
distinct priming parameters. Devices are primed consecutively, beginning with the
syringe dispensers, not concurrently.
Important
: Remove the Peri-pump cassette when the Syringe dispenser is set to
"Submerge."
"Submerge" is not offered for the Peri-pump because it is not a good practice for
the dispense cassettes. However, because the cassette Tip Holder resides on the
same dispense arm as the Syringe dispenser manifold, it will be moved into the
submerge position when the Syringe tips are submerged. Either remove the cassette
altogether when soaking the Syringe tips, or remove the prime trough inserts for
BioTek Instruments, Inc.