Perform the Syringe Dispense Precision & Accuracy Test | 236
Syringe Dispenser Test Materials
l
Microplates: 384-well plates for testing the 16-tube dispensers (which can be replaced
with 96-well plates if more applicable for your lab); 1536-well plates for testing the 32-
tube dispensers; and 96-well plates for testing the 8-tube dispensers. For the low-
density-plate-type specific manifolds, the applicable plate type is required.
l
Precision balance with readability of 0.0001 g resolution is preferable, 0.001 g resolution
is acceptable, and capacity of 100 g minimum
l
Pipettes and graduated beakers
l
Microplate absorbance reader capable of dual wavelength reading at 630 and 405 (or
450) nm
l
The test solutions:
Syringe Dispenser Test Solutions below
n
Important Recommendations for All Liquid Tests on page
Syringe Dispenser Test Solutions
20 µL Solution
o
Mix 10 mL of BioTek’s blue dye solution with 100 mL of deionized or distilled water
to create a dilution of the concentrate. Mix 160 mL of deionized or distilled water
with 20 mL of the diluted concentrate.
80 µL Solution
o
Mix 120 mL of deionized or distilled water with 40 mL of the
20 µL Solution
.
1536 Solution
o
Mix 21 mL of 70% Isopropyl Alcohol with 13 mL of the
20 µL Solution
and 150 mL
of DI H2O.
Perform the Syringe Dispense Precision & Accuracy Test
Prerequisites:
l
Gather the required materials:
Syringe Dispenser Materials above
l
Prepare the test solutions:
Syringe Dispenser Test Solutions above
.
l
Make a copy of the applicable worksheets.
BioTek Instruments, Inc.