Dispense Precision and Accuracy Test Worksheet
12-Well-Plate Syringe Manifold
Serial Number:
Dispense Precision Test
Standard Deviation:
(calculate using spreadsheet program)
Mean OD:
(sum of all wells
÷
12 wells)
% Coefficient of Variation:
((Standard Deviation
÷
Mean OD) x 100)
% CV <= 4.0?
q
Pass
q
Fail
Dispense Accuracy Test
Total Dispense (Actual) Weight:
grams
Expected Weight:
(µL/well x number of wells dispensed)
grams
% Accuracy Error:
(Actual Weight – Expected Weight) ÷ Expected Weight x 100
%
% Accuracy Error must be ≤ ±1.0%
c
Pass
c
Fail
Visual verification that no well varies
considerably from the others
c
Pass
c
Fail
Date:
Test Performed By:
MultiFlo FX Qualification Procedures
© BioTek® Instruments, Inc.