Peri-pump Dispense Precision & Accuracy Test Worksheet
10 µL Test for 10 µL Cassette
10 µL Dispense Precision Test
Standard Deviation (SD):
Mean Absorbance (sum of all wells ÷ # of wells)
% CV (SD ÷ Mean x 100)
%
% CV must be < 5.0%
c
Pass
c
Fail
10 µL Dispense Accuracy Test
Total Dispense (Actual) Weight:
grams
Expected Weight:
(mL/well x number of wells dispensed)
grams
% Accuracy Error:
(Actual Weight – Expected Weight) ÷ Expected Weight x 100
%
% Accuracy Error must be < 4.0%
c
Pass
c
Fail
Visual verification that no well varies
considerably from the others
c
Pass
c
Fail
Cassette Serial Num-
ber:
Tests Performed By:
Date:
Reviewed/Approved By:
Date:
MultiFlo FX Qualification Procedures