Replace Components | 211
Recalibrate the Peri-pump Dispense Cassette
Gravimetric Method
The most precise method for calibrating a cassette is the gravimetric method. Find
the instructions and an Excel
®
spreadsheet on the MultiFlo FX Operator’s Manual
USB stick, titled:
Gravimetric Calibration Instructions 7171024 (
current Rev
.
Calibration Jig
BioTek offers an accessory for recalibrating dispense cassettes. The Calibration Kit
(PN 7170017) includes an 8-pipette jig, instructions and an Excel spreadsheet that
may save time when calibrating a cassette.
Follow the instructions shipped with the kit or find them on the MultiFlo FX
Operator’s Manual USB stick, titled:
Calibration Jig Instructions 7171009 (
current
Rev
.
Important:
After calibrating the cassette, BioTek recommends performing the "Peri-
Pump Dispense Precision and Accuracy Tests" defined in the operator's manual and
IQOQPQ procedure to verify the cassette meets specifications.
Replace Syringe Dispenser Check Valve
You can order replacement check valves from BioTek Customer Care if your check
valves become clogged and cleaning them does not solve the problem:
l
PN 68083 — Autoclavable valves for use with non-organic substances.
l
PN 68073 — Check valves recommended for use with organic substances.
n
If you observe a decline in performance after changing the check valves,
Calibrate the Backlash for Syringe Dispenser on the next page
.