Maintenance
3500/42 Operation and Maintenance
170
Acceleration Test Setup
The Test Equipment outputs should be floating relative to earth ground.
Verification Screen Setup - Acceleration
Run the Rack Configuration Software on the test computer. Choose Verification
from the Utilities menu and choose the proper Slot number and Channel number
then click on the Verify button.
Note
If the Timed OK Channel Defeat is enabled, it will take 30 seconds for a
channel to return to the OK status from not OK. If OK MODE is configured for
latching, press the RESET button on the Rack Interface Module (RIM) to return
to the OK status.
Multimeter
Function Generator
Power Supply