Maintenance
3500/42 Operation and Maintenance
96
5.1.5
Radial Vibration Channels
The following sections describe how to test alarms, verify channels, and test OK
limits for channels configured as Radial Vibration. The output values and alarm
setpoints are verified by varying the input vibration signal level (both peak to
peak amplitude and DC voltage bias) and observing that the correct results are
reported in the Verification screen on the test computer.
Radial Vibration channels can be configured for the following channel values and
alarms:
Channel Values
Alarms
Over
Under
Direct
X
Gap
X
X
1X Amplitude and Phase
X
X
2X Amplitude and Phase
X
X
Not 1X Amplitude
X
S
max
Amplitude
X
5.1.5.1
Test Equipment and Software Setup - Radial Vibration
The following test equipment and software setup can be used as the initial set up
needed for all the Radial Vibration channel verification procedures (Test Alarms,
Verify Channels, and Test OK Limits).
Application Alert
WARNING
Tests will exceed alarm
setpoint levels causing
alarms to activate. This
could result in a relay
contact state change.
High voltage present.
Contact could cause
shock, burns, or death.
Do not touch exposed
wires or terminals.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com