3500/42 Operation and Maintenance
Maintenance
177
Full Scale Formulas - Integration
(For the Following units: in/s pk, in/s rms, mm/s pk, mm/s rms)
To input rms volts for peak full scale units:
To input rms volts for rms full scale units:
To input peak to peak volts for peak full scale units:
To input
peak to
peak volts
for RMS
full scale
units:
Input
Voltage
(V rms)
=
Full - scale (English units)
30.72
Scale Factor
(English units
0.1 volts / g typical)
Velocity Frequency
x 0.3535
/
Input
Voltage
(V rms)
=
Full - scale (English units)
30.72
Scale Factor
(English units
0.1 volts / g typical)
Velocity Frequency
x 0.5
/
Input
Voltage
(V pp)
=
Full - scale (English units)
30.72
Scale Factor
(English units
0.1 volts / g typical)
Velocity Frequency
x 1.414
/
Input
Voltage
(V pp)
=
Full - scale (English units)
30.72
Scale Factor
(English units
0.1 volts / g typical)
Velocity Frequency
/
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com