Panel Lockout
4-66
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Panel Lockout
Activating/Deactivating Panel Lockout
The Panel Lockout feature helps minimize the potential for keypad
tampering. It disables all the front panel keys except the
Main Display
,
Volume History
,
Channel Select
(for triple channel pumps only),
and
Back Light
keys, and the
Options
,
Primary
, and
Secondary
soft keys
for viewing.
Panel Lockout can only be enabled from the Main Display or the
Programming screen when the following conditions are met:
no alarms or alerts are present
for single channel pumps: an infusion is running
for triple channel pumps: all infusions are running, or up to two
infusions are on Standby. Panel Lockout cannot be enabled if all
three pump modules are on Standby.
To enable the panel lockout function, press the PANEL LOCKOUT
button located on the back of the pump. The Lock icon is displayed
between the second and third soft keys on the main display (Figure
4-95).
If disabled keys are pressed when the keypad is locked, the Panel
Locked pop-up is displayed (Figure 4-96).
To unlock the front panel, press the PANEL LOCKOUT button again.
Figure 4-95 Panel Locked with
Primary Running, No Alarm or Alert
Figure 4-96 Front Panel Locked
Invalid Key Pressed
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...