COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
4-39
Operating Instructions
Primary Infusions
12.
Verify that all values are appropriate and:
that pump programming matches the label on the source
container and the physician’s order
that the loaded administration set is connected to the correct
source container and administration route for the programmed
infusion
13.
Press the
Confirm Primary
soft key.
14.
Press the appropriate
START
key to begin the infusion. The Main
Display screen shows the rate, volume remaining, and the dose
(Figure 4-57).
If a secondary infusion has been programmed but not started, a
pop-up appears (Figure 4-26) warning that a primary infusion is
about to start when a secondary infusion has been programmed. See
step 7 on page 4-22 for more information.
Figure 4-56 Unconfirmed Primary
Program
Note:
Unconfirmed
Primary Program
If the
START
key is pressed before pressing the
Confirm Primary
soft key to verify the parameters, an
Unconfirmed Primary Program
alarm occurs and a
message is displayed on the prompt line (Figure 4-56).
Figure 4-57 Main Display
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...