8
Troubleshooting Alarms
8-14
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
NO BATTERY - Plug in Now
BATTERY/DEPLETED
The batteries are depleted and
infusions have stopped. The
pump must be plugged into AC
power before infusions may be
restarted. After 5 minutes in this
alarm state (Figure 8-14), the
pump will shut down.
To clear the alarm, plug into an AC power
source immediately. A pop-up is displayed,
instructing not to unplug the pump. Press the
Ok
soft key to clear the pop-up and resume the
infusion using AC power.
• To resume infusions on single channel pumps:
press the
START
key.
• To resume infusions on triple channel pumps:
press the
Channel Select
key(s), then press
the
START
key.
Do not use the pump on battery power until the
battery charge icon indicates that the batteries
have fully charged
If infusions are complete, power off the pump
by pressing the
ON/OFF CHARGE
key twice
and allow the batteries to recharge fully.
See “Battery Charge Alerts and Alarms,” 4-10
for more information.
Panel Lockout Button Stuck
STK KEY
(on all channels)
Without an infusion
programmed, the
PANEL
LOCKOUT
button has been
pressed for longer than
5 seconds.
Do one of the following as appropriate:
• Release the button.
• Relocate any objects next to the pump that may
be pressing the button.
• Ensure the pump’s power cord is not wrapped
around the pump in such a way as to press the
button.
If unresolved for more than 50 seconds, a
failure will result.
X - Primary Out of Range
X - Secondary Out of
Range
STOPPED
A programming value outside
the programmable range allowed
has been entered. This alarm
occurs as soon as the
Confirm
Primary
or
Confirm
Secondary
soft key or the
START
key is pressed.
• Verify the appropriate values have been
entered.
• Press the
Confirm Primary
or
Confirm
Secondary
soft key as appropriate, if
required, and the
START
key to begin the
infusion.
• If alarm recurs, the value range available in the
current PERSONALITY feature set may not be
broad enough to accommodate the entries.
Using the Options menu, check the current
PERSONALITY feature set configuration.
X - Reset Manual Tube
Release
RESET/label
The Manual Tube Release was
activated.
Close the regulating clamp on the
administration set,
remove the administration
set, and reset the Manual Tube Release. See
“Using the Manual Tube Release,” 4-71 for
details.
Table 8-1
Troubleshooting Alarm Messages — continued
Alarm Message
Cause
Recommended Action
Figure 8-14 NO BATTERY
Alarm Pop-up
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...